DRK506 Particulae Filtration Efficens Detector (PFE) -EN149

Description:

Introduction】 Persona medica chirurgica, particula filtration efficientiae PFE probator, aptus ad inspectionem centrum instrumenti medici, centrum inspectionis praesidium tutelae, centrum inspectionis medicinae tutelae, centrum inspectionis medicamentorum, centrum morbi praeventionis et temperantiae, inspectio centri textilis, valetudinarium, larva et fabricatores respiratoris; etc. Features PFE testor ad efficientiam particulae filtration larvarum chirurgicarum instructa est speciali generante sal aerosol, quod producere potest...


  • MALITIA Price:US $0.5 - 9,999 / Piece
  • Min.Order Quantitas:100 Piece/Picies
  • Facultates copia:(X) Piece / Mass per mense
  • Portus:Shenzhen
  • Pensio conditio:L/C,D/A,D/P,T/T
  • Product Detail

    Product Tags

    .Introductio.

    Persona medica chirurgica particula filtrationis efficientiae PFE probator aptus est ad centrum inspectionis instrumenti medici, centrum inspectionis tutelae securitatis, centrum inspectionis laboris tutelae, centrum inspectionis medicamentorum, centrum morbi praeventionis, centrum inspectionis textilis, hospitalis, larvae et fabricatores respiratoris, etc.

    .Features.

    PFE probator pro particula filtrationis efficientia larvarum chirurgicarum instructa est speciali generante sal aerosol, quod potest producere aerosoles cum particula speciei et intentionis specificae.

    Instructus serie speciali fixture, omni genere larvarum detectioni apta.

    Vita in alta vita modulus photometer constructus, temporis cumulus sampling, lux purgatio prompta.

    Automatic imperium generationis aerosol, calculi automatismi capiendi efficientiam et resistentiam airflow larva, interventus minuere.

    PC imperium cum programmate Drick, facile ad operandum et famam probationis servabit!

    Instructa in alta praecisione fluxus electronici electronicus et summus effectus sampling sentinam ad stabilitatem fluunt curent.

    .Standard .

    EN149: Respriatorium machinas tutelae medias personas percolando tueri contra particulas-Requisita, probatio, notatio

    EN13274-7: Respriatorium machinas tutelae-Modos probatoris - Pars VII: Determinatio particulae sparguntur penetrationem
    .Technical Parameters.

    Item Detail
    Particula salis CMD 0.093µm
    Particula olei CMD 0.185μm
    Influunt rate 5-100 L/min±1%
    Deprehensio particula range (0.001-100) μg/L
    Test pressura range 0-250Pa ±2.0%
    Test aerosol DEHS, oleum Paraffin, NaCl
    Test area 100cm2
    Particula numerus Photometer
    Ostendere X "Tange screen imperium panel"
    Data export Typographus embedded
    Potentia copia 220V 50-60 Hz
    Dimensiones 91X80X173cm

    .Main configurationis Photos.

     

     

     pro_imgs4

     

     pro_imgs5

     pro_imgs6

    PFE Tester Pharusparticula contra PC Cum Software

     

     

     pro_imgs7

     

     pro_imgs8

     pro_imgs9

    Aliquam pila particula

    "Parum" ex US

    N95 Face larva Fixture


  • Previous:
  • Deinde:

  • Related Products

    Whatsapp Online Chat!