DRK-1000 Bacterial Filtration Efficiency Detector

Short Description:

Product introduction DRK-1000 Bacterial Filtration Efficiency Detector not only conforms to the requirements of medical surgical masks technology YY0469-2011 appendix B in the test method for filter efficiency (BFE) bacteria first b. 1.1.1 test instrument, but also conforms to the American society for testing material ASTMF2100, ASTMF2101, the requirements of the European EN14683 standards, innovative improvements were made on the basis of this, with double pneumatic contrast sampling method ...


  • FOB Price: US $0.5 - 9,999 / Piece
  • Min.Order Quantity: 100 Piece/Pieces
  • Supply Ability: 10000 Piece/Pieces per Month
  • Port: Shenzhen
  • Payment Terms: L/C,D/A,D/P,T/T
  • Product Detail

    Product Tags

    Product introduction

    DRK-1000 Bacterial Filtration Efficiency Detector not only conforms to the requirements of medical surgical masks technology YY0469-2011 appendix B in the test method for filter efficiency (BFE) bacteria first b. 1.1.1 test instrument, but also conforms to the American society for testing material ASTMF2100, ASTMF2101, the requirements of the European EN14683 standards, innovative improvements were made on the basis of this, with double pneumatic contrast sampling method at the same time, improve the precision of sampling,It is suitable for measuring and testing departments, scientific research institutes, mask manufacturers and other relevant departments to test the performance of mask bacterial filtration efficiency.

    Execution standard

    YY0469

    Q/0212 ZRB003-2015 

    ISO/DIS 22611

    ASTMF2101

    EN14683

    Product features

    1.Negative pressure test system to ensure the safety of operators;

    2.Built – in peristaltic pump in negative pressure cabinet, A, B two – way, six Andersen;The peristaltic pump flow can be set;

    3.The flow rate of night spray can be set and the atomization effect is good.

    4.Embedded high speed industrial microcomputer control;10.4 inches industrial high brightness color touch screen;

    5.USB interface, support U disk data storage;

    6.Cabinet built-in high brightness lights;Built-in leakage protection switch to protect the safety of operators;

    7.The inner layer of the cabinet is made of stainless steel, the outer layer is sprayed with plastic and cold-rolled, and the inner and outer layers are insulated and flame retardant. 

    8.The front switch type glass door is convenient for the experimenter to observe and operate.

    9.Detachable bracket, adjustable bracket height;Supporting and moving dual purpose casters.

    Specifications:

    Key Specification

     

    Specification Range

    Resolution

    Accuracy

     

    A Route Sampling Flow

    28.3L/min

    0.1L/min

    Within ± 2.5%

    B Route Sampling Flow

    28.3L/min

    0.1L/min

    Within ± 2.5%

    Spray Flow

    (8~10)L/min

    0.1L/min

    Within ± 2.5%

    Peristaltic Pump Flow

    (0.006~3.0)mL/min

    0.001ml/min

    Within ± 2.5%

    Front Pressure of A Route

    Sampling Flowmeter

    (-20~0)kPa

    0.01kPa

    Within ± 2.5%

    Front Pressure of B Route

    Sampling Flowmete

    (-20~0)kPa

    0.01kPa

    Within ± 2.5%

    Front Pressure of

    Spray Flowmeter

    (0~300)kPa

    0.01kPa

    Within ± 2.5%

    Ambient Temperature

    (-40~99)℃

    0.1℃

    Within ± 2.5%

    Negative Pressure of the Aerosol

    Chamber

    (-90~-120)Pa

    0.1Pa

    Within ±2.0%

    Negative Pressure of the Cabinet

    -50~-200pa

     

    Data Storage Ability

    >100000 sets

     

    Vortex Mixer Test Tube

    Specification and Quantity

    Φ16×150mm test tube,8pcs

    High Efficiency Particulate Air

    Filter Properties

    Filter Efficiency≥99.99% for particles> 0.3μm

    Median diameter of aerosol  generator mass

    Average diameter: (3.0±0.3)μm; Geometric Standard

    generator mass Deviation≤1.5

     

    Double Routes 6-lever Andersen Sampling apparatus

    Level1>7μm; Level 2: (4.7~7)μm;

    Level 3: (3.3~4.7)μm; Level 4: (2.1~3.3)μm;

    Level 5: (1.1~2.1)μm; Level 6: (0.6~1.1)μm

    Size of Aerosol Chamber

     

    600×85×3mm (Length×Diameter×Thickness)

     

    Total number of positive quality

    control sampler particles

    (2200±500)cfu

     

    Negative Chamber Flow Speed

    ≥5m3/min

     

    Size of Negative Pressure

    Chamber Door(W×D)

    1000×730mm

    Size of Main Machine(W×D×H)

    1180×650×1300mm

     

    Size of Supporting

    Frame(W×D×H)

    1180×650×600mm, Height within 10cm is adjustable

    Noise of the Tester

    <65dB(A)

     

    Weight

    150kg

     

    Power Supply

    AC220V±10%, 50Hz

     

    Power Consumption

    <1500W

     


  • Previous:
  • Next:

  • Related Products

    WhatsApp Online Chat !