DRK703 Mask field tester

Short Description:

Main Uses According to relevant standards, it is mainly used for visual field testing of masks, masks and respirator, which is suitable for mask and respirator manufacturers, quality supervision, scientific research, wearing and using units. Main features 1. The whole equipment is composed of semi-arc bow, recording device, seat frame and test head die. 2. Semi-arc bow: the radius (300-340) mm can be rotated around the horizontal radius plane passing through the point 0°, and a scale extends ...


  • FOB Price: US $0.5 - 9,999 / Piece
  • Min.Order Quantity: 100 Piece/Pieces
  • Supply Ability: 10000 Piece/Pieces per Month
  • Port: Shenzhen
  • Payment Terms: L/C,D/A,D/P,T/T
  • Product Detail

    Product Tags

    Main Uses

    According to relevant standards, it is mainly used for visual field testing of masks, masks and respirator, which is suitable for mask and respirator manufacturers, quality supervision, scientific research, wearing and using units.

    Main features

    1. The whole equipment is composed of semi-arc bow, recording device, seat frame and test head die.
    2. Semi-arc bow: the radius (300-340) mm can be rotated around the horizontal radius plane passing through the point 0°, and a scale extends to the 90° arc bow every 5° from 0° on both sides, equipped with sliding white visual standard.
    3. Recording device: the recording needle is connected with the visual standard through components such as axle wheel, and the direction and Angle of the visual standard are recorded on the visual field drawing correspondingly.
    4. Seat frame: used to support semi-circular arc bow and fixed recording device.
    5. Test the head mold: a standard head mold, the two eye holes of the head mold are equipped with small light bulbs, the position of the bulb and the position of the head mold are according to the requirements of GB 2890. The test head die shall be installed on the workbench in such a way that the left and right eyes are placed at the center of the semicircular arch and the “0″ point is directly seen.

    Main indicators

    1. Semi-circular arc radius: 335mm.
    2. Left and right field of view: ≤120°.
    3. Bulb distance: the vertex of the bulb is connected after the two eye points (7±0.5).

    Applicable standards

    GB/T 32610-2016, GB 2626-2019, gb2890-2009